- Ships in 2 Weeks, FREE SHIPPING!
- Compact
- Inexpensive
- Sensor With Long Operating Life
- Accuracy: +/- 2% FSD T = const. & +/-5% FSD 0>T.50C (ppm)
- Reverse Voltage Protection & Temperature Compensation
- Application: Air, Oxygen and Nitrogen Generators, Heat Treating, Bright Annealing, Gas Mfgs, Filling Stations
- One User Specific Measuring Range (0-10%, 0-25% , 0-100%)
The O2TRACER-DIS-R series are remote configured transmitters that are accurate and very simple to use in the measurement of oxygen in the % range. The units are offered in three standards % ranges (0-10%, 0-25% , 0-100%).
The units use a special fuel cell to measure he oxygen concentration. The sensor meets the industrial requirements for accuracy, sensitivity and is easy to use and has a long operating life.
Calibration of the instrument for trace oxygen measurements in gas should be done with a calibration gas. The concentration can be chose freely within the measuring range. In the % range the unit can be calibrated with air.
To Order the models below scroll to the top of the page and choose the desired model from the "RANGES" drop down menu, then ADD TO CART |
PART NUMBER | RANGE |
OX-RO-O2T-DIS-R1 | 0-10% |
OX-RO-O2T-DIS-R2 | 0-25% |
OX-RO-O2T-DIS-R3 | 0-100% |
SPECIFICATIONS
MEASURING RANGES: 0-10%, 0-25%, 0-100%
CALIBRATION: with calibration gas (ppm) or air
ACCURACY: +/- 2% FSD T = const. & +/-5% FSD 0>T.50C (ppm)
RESPONSE TIME: 90% FSD @ 25C, 0-100% < 10 s
OPERATING TEMPERATURE: 0-50C
PRESSURE: 0.1 - 1 bar
SIGNAL OUTPUT: 4 - 20 mA/DC
ALARM VALUE: 22.8 mA (standard) or 3.4 mA (if desired)
VOLTAGE: 10 -35 Vdc, reverse voltage protection up to 40 Vdc
LOAD: typ. 470 ohm, max 750 ohm
DISPLAY: 6 digit display with backlight
OXYGEN SENSOR: micro-fuel cell
HOUSING: IP65
SIZE: 80 x 75 x 55 mm
WEIGHT: 0.3 kg
ENCLOSURE: aluminum
CONNECTION: 6 mm or 1/4 tubing