N1500-G Universal Input Large Display Panel Meter with 56mm High Digit

Free Shipping! - USA Orders Over $75

Customer Reviews
N1500-G Universal Input Large Display Panel Meter with 56mm High Digits

N1500-G Universal Input Large Display Panel Meter with 56mm High Digits

Regular price $444.00 Sale SKU:N1500-G-RS485
1 In stock* Ships in 3-4 weeks, if NOT in stock
* ships same or next day

  • Accepts Thermocouples J, K, T, E, N, R, S, B; Pt100, 4-20 Ma, 0-50 mV And 0-5 Vdc
  • Programmable Range From -1999 To +9999 With Adjustable Filter
  • Alarms: Two 3 A SPST Relays
  • Functions: Lo, Hi, Differential, Diff. Lo, Diff. Hi And Sensor Break
  • Hold, Peak, Max And Min Function
  • 24 Vdc Output For Remote Transmitters
  • RS485 Modbus Digital Communication

The N1500-G is a high performance large display indicator which can accept inputs from most sensor types including thermocouples, RTDs, voltage and current.
Its large 56mm LED digits make the N1500-G a great choice when you need a display which is readable from a distance.
The N1500-G also includes two 3 A SPST relay which can be used for alarm or controlling purposes.  It also has RS485 communications which allows the N1500-G to transmit data to a computer, PLC or SCADA system.

 To Order the model below scroll to the top of the page then  click on ADD TO CART button
 PART NUMBERS DESCRIPTION
N1500-G-RS485 Universal Input Large Display Panel Meter with 56mm High Digits

 

 Downloads

User Manual

N1500G Registers for Modbus Communication

 

 

 

SPECIFICATIONS

Ranges
J Thermocouple: -50 to 760 °C (-58 to 1400 °F)
K Thermocouple: -90 to 1370 °C (-130 to 2498 °F)
T Thermocouple: -100 to 400 °C (-148 to 752 °F)
E Thermocouple: -35 to 720 °C (-31 to 1328 °F)
N Thermocouple: -90 to 1300 °C (-130 to 2372 °F)
R Thermocouple: 0 to 1760 °C (-32 to 3200 °F)
S Thermocouple: 0 to 1760 °C (-32 to 3200 °F)
B Thermocouple: 150 to 1820 °C (302 to 3308 °F)
RTD (Pt100): -199.9 to 530.0 °C (-328 to 986.0 °F)
Current: 4-20 mA, 0-50 mV
Voltage: 0-5 V

Internal resolution: 19500 levels, display: 12000 levels (-1999 to 9999)

Accuracy
Thermocouples J, K, T and N: 0.25% of span ±1°C.
Thermocouple E, R, S and B: 0.25 % of span ±3°C.
Pt100: 0.2 % of span
Current or linear voltage: 0.2% of the maximum range

Pt100 measurement: 3 wire circuit. Current 0.170 mA.

Input impedance
 0-50 mV, Pt100 and thermocouples: >10 MΩ
 0-5 V: >1 MΩ
 0-20 mA, 4-20 mA: 22 Ω

Relays: Two SPST  3A / 250 Vac
Auxiliary Voltage Source: 24 Vdc  (±10 %) / 25 mA max.
Operating temperature: 0 to 55 °C, Maximum RH: 80 % up to 30 °C. For temperatures above 30 °C, decrease 3 % per °C
Approximate weight: 1 kg
Dimensions: 310 x 110 x 37 mm (C x A x P)
Power: 100 to 240 Vac/dc ±10 %; 50/60 Hz
Communications:  RS485, Modbus protocol
Max. Consumption: 10 VA